Radiation Exposure and Image Quality Statistics for All Facilities Performing AP Lumbar Spine Radiographs

Radiation Exposure

(Before QA (year 0) and QA (years 7-16)

Machines that measured high or extremely high patient radiation exposure decreased from 47% in the first year to 5% in year-sixteen of the quality assurance program. Average radiation exposure has decrease by 46% for lumbar spine examination since the implementation of QA.
Image quality scores are measured and reported on a scale of zero to one hundred. Image quality scores are determined by the inspector’s evaluation of an x-ray taken of an image quality tool and processed using the facility’s processing equipment. The total image quality score is compiled by the inspector evaluating six areas of image quality including background density, high contrast resolution, low contrast resolution, density uniformity and noise/artifacts. Image quality scores of 0-29 are reported as poor; 30-49 as fair; 50-69 as good and 70 or greater as excellent image quality. Image quality scores have dramatically improved from year-one to year-sixteen of the quality assurance program. In year-one, 52% of machines evaluated had fair or good image quality scores (ie./ scores of 49 or lower). In year-sixteen, 2% of all machines evaluated had image quality scores or 49 or lower.

Return to QA Page

AQES: Home
Department: NJDEP Home | About DEP | Index by Topic | Programs/Units | DEP Online
Statewide: NJ Home | Services A to Z | Departments/Agencies | FAQs

Copyright © State of New Jersey, 1996-2024

Last Updated: October 27, 2021